ReviewData extraction and reporting strategies of studies assessing non-central corneal thickness by Pentacam: A review
Introduction
Non-invasive Scheimpflug imaging systems such as the Pentacam (Oculus GmbH, Wetzlar, Germany) provide global pachymetry maps, i.e. corneal thickness (CT) measurements along various corneal meridians. Central (CCT) and non-central corneal thickness (NCCT) are diagnostic parameters used to detect sub-clinical keratoconus in preoperative refractive surgery assessments [1]. NCCT has been considered in corneal crosslinking [2], [3] and may provide useful information for corneal surgery at non-central locations such as intra-stromal ring insertion and limbal relaxing incisions [4].
The Pentacam is a non-contact device that uses a monochromatic blue light (475 nm) and a rotating Scheimpflug camera, which revolves 360 degrees around the optical axis, to scan the anterior segment and its main features have been described previously [5]. In brief, either 25 or 50 images can be obtained per scan and the procedure takes up to 2 s. Prior to image acquisition, the patient is asked to place their chin onto a chinrest and their forehead against a bar and to fixate on a small round target that is located at the centre of the vertical blue light slit. The instrument is centred on the pupil, using real time monitoring on the instrument's screen. A second, camera is used to monitor fixation. The scan is usually initiated automatically, but manual acquisition can be selected. Following image acquisition, the observer can view a wide range of topographic maps. A detailed pachymetric map can be obtained by selecting the appropriate option on the Pentacam system. The pachymetry readings are calculated from the spatial difference between anterior and posterior corneal surface, usually perpendicular to the anterior corneal surface (normal-to-surface tangent measurement, personal communication with manufacturer) [6]. These calculations could be considered to approximate radial pachymetry measurements. The maps contain up to 81 single-point numerical pachymetry readings (Fig. 1) and further data points can be obtained by moving the mouse-based cursor over the map. Visual inspection of the colour-coded maps provides a gross overview of any central and non-central abnormalities. Additional features of the Pentacam software include pachymetry at 3 mm from the centre (Fig. 2) and the illustration of the averaged corneal thickness spatial profile as well as the percentage thickness increase from the centre to the periphery near the limbus, both of which are based on pachymetry readings from ring-averaged thickness along 22 imaginary circles that are centred on the thinnest point (Fig. 3, Oculus Pentacam Interpretation Guideline, 2nd ed) [1]. An advantage of the Pentacam over similar systems is that it can provide non-central pachymetry data up to 5 mm away from the corneal centre, whereas earlier corneal imaging instruments, such as Orbscan, routinely provide data up to 4 and occasionally 4.5 mm away from the centre. Three versions of the Pentacam are currently in clinical use, the Pentacam Basic, Classic, and HR (high resolution). The basic and classic models differ mainly in their software features and the HR model also provides additional hardware updates [5].
While non-contact computerized pachymetry systems are advantageous in reducing inter-observer measurement variability, the complexity of the pachymetry information may present observers with challenges related to the question as to which data should be extracted and compared. For example, observers have the choice between points (Fig. 1, Fig. 2) versus ring-averaged pachymetry (Fig. 3) and will also have to decide from which geographic location to obtain the measurements. The evaluation of peer-reviewed Pentacam studies that report on NCCT confirms that authors have indeed adopted a number of different strategies to extract and report non-central pachymetry measurements. This can be problematic as the use of different data extraction strategies could potentially obscure physiological variations such as changes in pupil size affecting the measurement [7]. Using ring-averaged pachymetry may lead to unexpected results in corneas of generally normal configuration and obscure regional variation in CT. The differences in data extraction and reporting not only make comparisons more difficult, but could potentially affect clinical decision-making, e.g. when assessing the effect of corneal crosslinking outside the corneal centre [2], [3]. To date, there has been little discussion regarding different data extraction strategies. The most commonly used methods in previous studies are single-point and ring-averaged data extraction (see below). Some researchers have provided pachymetry for up to 22 0.4 mm increments from the centre [1], others have reported CT measured at a single distance in close proximity to the central cornea, e.g. 1.5 mm away from the centre [8].
The purpose of this paper is to review data extraction strategies used in different studies to report non-central corneal thickness measurements, with a focus on single-point and ring-averaged data obtained with the Pentacam system.
Section snippets
Materials and methods
A literature search was carried out on 15 September 2013 using the United States National Library of Medicine PubMed database. Pre-defined search terms were used individually as well as in combination. The search terms included corneal thickness, peripheral corneal thickness, off-centre corneal thickness, non-central corneal thickness, pachymetry, Pentacam, and Scheimpflug imaging. The abstracts of studies with relevant titles were reviewed to ascertain relevance. All studies with relevant
Results
Seventeen relevant Pentacam studies published between 2005 and 2013 were identified. Combined, these studies assessed at least 3008 eyes of 2485 subjects, the age of subjects ranged from 7 to 83 years. The studies included individuals of various ethnic groups and were carried out in Australia [7], [9], [10], Austria [8], Brazil [1], China [11], [12], [13], France [14], Germany [15], [16], Iran [17], Malaysia [18], Mexico/USA [19], Spain [20], and the UK [4], [21]. Non-central pachymetry
Discussion
This review discusses approaches that have been adopted for the reporting of non-central pachymetry obtained with the Pentacam system. A main difference between studies was the use of either point-pachymetry from a defined location and ring-averaged pachymetry.
Funding
This work was supported by Santander Universities Fund UK.
Conflict of interest
The author declares no conflict of interest.
Acknowledgements
The author would like to thank the reviewers for their constructive comments and support in improving the manuscript. The author also thanks Peter Rafferty, Graeme Kennedy, Andreas Steinmueller, and Sven Reisdorf for helpful discussions on optical aspects of the Pentacam.
References (29)
- et al.
Corneal-thickness spatial profile and corneal-volume distribution: tomographic indices to detect keratoconus
J Cataract Refract Surg
(2006) - et al.
Correlation of central and peripheral corneal thickness in healthy corneas
Cont Lens Anterior Eye
(2012) - et al.
Corneal pachymetry: new ways to look at an old measurement
J Cataract Refract Surg
(2014) - et al.
Anterior segment biometry with the Pentacam: comprehensive assessment of repeatability of automated measurements
J Cataract Refract Surg
(2008) - et al.
Corneal changes following short-term rigid contact lens wear
Cont Lens Anterior Eye
(2012) - et al.
A comparison between Scheimpflug imaging and optical coherence tomography in measuring corneal thickness
Ophthalmology
(2013) - et al.
Average 3-dimensional models for the comparison of Orbscan II and Pentacam pachymetry maps in normal corneas
Ophthalmology
(2009) - et al.
Central and peripheral pachymetry measurements according to age using the Pentacam rotating Scheimpflug camera
J Cataract Refract Surg
(2007) - et al.
Central and peripheral corneal thickness measured with optical coherence tomography, Scheimpflug imaging, and ultrasound pachymetry in normal, keratoconus-suspect, and post-laser in situ keratomileusis eyes
J Cataract Refract Surg
(2009) - et al.
Fuchs’ endothelial corneal dystrophy
Ophthalmology
(2013)
Pentacam assessment of posterior lamellar grafts to explain hyperopization after Descemet's stripping automated endothelial keratoplasty
Ophthalmology
Spatial distribution of corneal light scattering after corneal collagen crosslinking
J Cataract Refract Surg
Riboflavin/ultraviolet a crosslinking of the paracentral cornea
Cornea
Corneal imaging with slit-scanning and Scheimpflug imaging techniques
Clin Exp Optom
Cited by (9)
The role of peripheral ocular length and peripheral corneal radius of curvature in determining refractive error
2022, Journal of OptometryCitation Excerpt :Corneal curvature measurements were obtained using Pentacam (Oculus GmbH, Wetzlar, Germany). Pentacam is a non-contact optical instrument with a Scheimpflug camera that rotates 360 degrees around the visual axis and scans the anterior segment of the eye within two seconds.19 Pentacam has shown good repeatability and reliability for central and peripheral corneal measurements.20,21
Corneal structure, transparency, thickness and optical density (densitometry), especially as relevant to contact lens wear—a review
2019, Contact Lens and Anterior EyeCitation Excerpt :For the Pentacam system, an optical section through the cornea also defines the anterior and posterior surfaces, with the instrument software applying a metric (shown as a slightly irregular red line in a colour image) to show the location of the anterior interface (Fig. 5C). From such locations, the instrument can be simply used to provide thickness values across the cornea [32]. It should be noted however that the light scatter from the anterior aspect of the cornea (which is the same viewed in the other figures) is more substantial, extending beyond where the actual corneal epithelium would be expected to be, i.e. is an anterior portion of the cornea comprising both epithelium and anterior stroma.
Photodynamic inactivation of multidrug-resistant Staphylococcus aureus by chlorin e6 and red light (λ = 670 nm)
2016, Journal of Photochemistry and Photobiology B: BiologyCitation Excerpt :These assays identified the corneal epithelium as a strong diffusion barrier for Ce6 when applied topically. Mechanical removal of the epithelium prior to the PS application, however, allowed the PS to deeply penetrate into the stroma, and yielded in Ce6 concentrations at a tissue depth of 1 mm much higher as those required to kill the pathogen in vitro, suggesting that PDT with Ce6 and red light might indeed be an promising therapy option to combat bacterial keratitis in humans caused by MDR-SA, especially when taking into account that the human cornea is only half as thick as its porcine analog [41,42]. To be effective as therapeutic agent, Ce6 needs to accumulate at/in the bacterial cells at doses much higher than those found at/in the surrounding tissue to selectively harm the invading pathogen without causing generalized tissue damage.
Using the posterior to anterior corneal curvature radii ratio to minimize the risk of a postoperative hyperopic shift after Descemet membrane endothelial keratoplasty
2020, Graefe's Archive for Clinical and Experimental OphthalmologyAccommodation response and spherical aberration during orthokeratology
2020, Graefe's Archive for Clinical and Experimental Ophthalmology