ORIGINAL ARTICLE
Influence of Fogging Lenses and Cycloplegia on Peripheral
Refraction
António Queirós, Jorge Jorge and José Manuel González-Méijome
ABSTRACT
PURPOSE: To compare objective peripheral refraction measured
with an open-field autorefractor without cycloplegia with the
values obtained with fogging lenses or with cycloplegia to inhibit
accommodation.
METHODS: For one hundred and sixty young adults aged 18 to 28
(mean 21.5 ± 2.3 years) their refraction was measured with the
Grand Seiko (GS) autorefractor at the center and at four peripheral
locations in the nasal and temporal directions under three different
conditions: 1) without cycloplegia (GS); 2) without cycloplegia,
but using a +2.00D fogging lens (GS_2D) and 3) with cycloplegia
(GS_cycl).
RESULTS: Mean spherical equivalent refraction (M) was significantly
more negative with the GS method in the hyperopic group
for central and peripheral refraction, and only at the center and at
10º nasal eccentricity for the emmetropic group (P<0.05, Kruskal-
Wallis). Paired comparison showed that differences of M values
across techniques were larger for the GS-vs.-GS_2D comparison in
myopes and emmetropes, and for the GS-vs.-GS_cycl one in hyperopes
(P<0.001, Wilcoxon Signed Ranks Test). The gap between M
values for all paired comparisons remained almost constant across
all eccentric positions under analysis.
CONCLUSIONS: Fogging lenses used with open-field autorefraction
up to 20º in the nasal and temporal fields seem to provide similar
accommodative relaxation to that provided by a cycloplegic. This is
particularly important when refracting emmetropes and hyperopes.
Moreover, this behavior seems to be independent of the eccentricity
at which measurements are taken.
(J Optom 2009;2:83-89 ©2009 Spanish Council of Optometry)
KEY WORDS: cycloplegia; fogging lenses; GrandSeiko; open-field
autorefraction; peripheral refraction.
RESUMEN
OBJETIVO: Comparar las medidas objetivas de refracción periférica
realizadas sin cicloplégico con los valores obtenidos con “lentes de
miopización” o con cicloplegia, ambas técnicas utilizadas para inhibir
la acomodación.
MÉTODOS: Se midió la refracción a 160 adultos jóvenes, con edades
comprendidas entre 18 y 28 años (media=21,5± 2,3 años), con un
autorrefractómetro Grand Seiko (GS), tanto en el centro del campo
visual como en 4 regiones de la periferia situadas nasal y temporalmente,
y todo ello en 3 condiciones diferentes: 1) sin cicloplegia
(GS); 2) sin cicloplegia , pero utilizando una lente translúcida de
+2.00 D (GS_2D) y 3) con cicloplegia (GS_cycl).
RESULTADOS: La media del equivalente esférico de la refracción
(M) resultó ser significativamente más negativa en la condición GS
en el grupo de los hipermétropes en lo que respecta a refracción
central y periférica, mientras que en el grupo de los emétropes sólo
ocurrió esto en el centro y a una excentricidad de 10º nasal (P<0,05,
Kruskal-Wallis). La comparación por pares de muestras relacionadas
reveló que la mayor diferencia de M entre distintas condiciones se
obtuvo al comparar GS y GS_2D en el grupo de los miopes y en
el de los emétropes, y al comparar GS y GS_cycl en el de los hipermétropes
(P<0,001, contraste de Wilcoxon de rangos con signo).
La discrepancia entre valores de M para las distintas comparaciones
por pares se mantiene prácticamente constante para todas las excentricidades
analizadas.
CONCLUSIONES: Las lentes de miopización, utilizadas con un autorrefractómetro
con ventana de observación para medir excentricidades
de hasta 20º en las direcciones nasal y temporal parece que
logran una relajación acomodativa similar a la que proporciona el
agente cicloplégico. Esto resulta especialmente relevante cuando se
evalúa la refracción en emétropes e hipermétropes. Además, este
comportamiento parece ser independiente de la excentricidad en la
que se realiza la medida.
(J Optom 2009;2:83-89 ©2009 Consejo General de Colegios de
Ópticos-Optometristas de España)
PALABRAS CLAVE: cicloplegia; lentes de miopización; Grand Seiko; autorefracción de campo abierto; refracción periférica.
From the Clinical and Experimental Optometry Research Lab. Department of Physics (Optometry), School of Sciences, University of Minho, Braga, Portugal.
Acknowledgements: We thank J. A. Díaz-Rey for his contributions in the
preliminary ophthalmological examination of patients and administration
of cycloplegic agent. None of the authors has a commercial or financial
interest in the instruments presented here.
Received: 4 March 2009
Revised: 2 April 2009
Accepted: 20 April 2009
Corresponding author: António Queirós
Department of Physics (Optometry). Campus de Gualtar.
University of Minho. 4710-057 Braga – Portugal
e-mail: aqp@fisica.uminho.pt.
The primary function of the peripheral retina is motion detection. This region is considered to have low ability for detail resolution, dominated by a considerable amount of astigmatism and field curvature.1
Peripheral refraction has been thoroughly studied over the last years, partly because myopia progression is considered to be associated to the hyperopic defocus of the images produced on the peripheral retina2,3 and to the potential influence on the emmetropization process of the human eye.2,4,5 On monkeys, it was possible to verify that the peripheral retina plays a role in myopia development during form deprivation.6-8
Even though early reports on peripheral refraction date back as early as the 1930’s,9 this subject has been more extensively studied since the 1970’s. The techniques of subjective refraction and retinoscopy were approached by Rempt et al., Millodot and Lamont and Wang et al.10-12 Peripheral refraction was also studied by means of objective and manual optometers;13,14 open-field autorefractors,1,16 photorefraction,4 double pass method,17 and clinical aberrometers.18 Several factors with the potential to affect peripheral refraction were also analyzed; namely age,2,19 refractive error,4,20 eccentric horizontal and vertical retinal locations,21 different fixation distance22 and the impact of anterior cornea reshaping with orthokeratology16 or LASIK.23
Several of the cited studies were intended to analyze peripheral refraction using a number of instruments,24 either under cycloplegia24 or without cycloplegic agents.21 Peripheral refraction has also been studied with the goal of obtaining a theoretical model of the ophthalmic lenses that could correct the peripheral refractive error.25
Refractive error in the peripheral retina, as compared to central refraction, shows a myopic trend in emmetropes and hyperopes. On the other hand, myopes present a peripheral refraction that is less myopic or more hyperopic than the central measurement;4,5,26 this fact has been more intensively studied for the horizontal meridian of the eye.21
The application of autorefractors to obtain peripheral refraction data raises the question of which is the effect of accommodation and which is the validity of such measures without cycloplegia. Other alternatives to the use of a cycloplegic, such as the use of fogging lenses, have demonstrated to be effective in evaluating central refraction.27 However, when it comes to peripheral refraction, the effectiveness of fogging lenses to avoid the accommodation effects in young adults being measured with an open-field autorefractor has not been investigated yet. The goal of this study is to evaluate whether or not peripheral measurements with an open-field autorefractor, up to 20º, are affected by the use of a fogging lens to inhibit the accommodative action, particularly in emmetropic and hyperopic young-adult populations.
METHODSThe population under test consisted of one hundred and sixty young university students, whose ages ranged from 18 to 28 years (mean ± SD: 21.5 ± 2.3 years). Of them, 114 were female (71.3%) and 46 male.
In order to separately analyze the potential effect of fogging
lenses and cycloplegia on accommodation during refraction,
three refractive groups were established according to
their spherical equivalent [M=Spherical+Cylinder/2, where
the values correspond to central refraction measurements
under cyclopegia], as follows: myopes (M≤-0.50; n=56;
35.0%), emmetropes (-0.50 After explaining them the nature of the study, each patient
signed a consent form before being enrolled. The research
followed the Declaration of Helsinki rules and was reviewed
and approved by the Scientific Committee of the School of
Sciences of Minho University (Portugal). The inclusion criteria
required that the subjects did not suffer from any current
eye disease or injury, were not taking any ocular or systemic
medication that could affect the accommodative response
and had best corrected visual acuity of 20/20 or better in
each eye. The intraocular pressure was checked with a noncontact
tonometer (Nidek Model NT-4000).30 The autorefraction data was obtained with an open-field
instrument, the Grand Seiko Auto Ref/Keratometer WAM-
5500 5500 (Grand Seiko Co., Ltd., Hiroshima, Japan). This instrument
had been used and tested by a number of authors to
measure refraction in the central and peripheral retina.19,31 The illumination of the room was adjusted so that the
pupil size was greater than 4mm, which was achieved in
all cases with and without cycloplegic agent. The fixation
target was placed at a distance of 5 meters from the patient’s
corneal vertex and consisted of a 5 LED arrangement in the
horizontal direction: a central one, two to its right and two
to its left. Each LED was separated from the adjacent ones
by an angular distance of 10º, measured at the patient’s
position. The subject was seated with the head stabilized by
a chin-rest so that the eye was aligned with the central LED.
For the right eye, the fixation of an object positioned at the
right hand of the central point (nasal visual field in the eye’s
primary position) matches the temporal retina measures. For
the right eye of each individual and for each retinal location
under analysis, three readings were taken and averaged, considering
the center of the pupil as the reference point of measurement.
The left eye was able to see the same target under
the same conditions (naked eye, with a +2.00 D fogging lens
or under cycloplegia). The refraction measurements were carried out in three
different conditions, and always in this sequence: 1) without
cycloplegia (GS); 2) without cycloplegia with the +2.00
D fogging lenses placed 12 mm away from corneal vertex
(GS_2D), and 3) with cycloplegia (GS_cycl). Following the
measurement with fogging lenses, cycloplegia was achieved
by instilling twice in each eye one drop of 1% cyclopentolate,
with a 5 minute interval between drops. After a 35 min wait
we were able to perform the measurements corresponding to
the cycloplegia condition. Participants were not wearing their correction; they were
only wearing the fogging spectacle lens (+2.00 D), consisting
of convex-concave (meniscus) lenses. The fogging lenses
were mounted on a trial frame and placed at a distance of 12
mm from the corneal vertex. The naso-pupilar distance was
adjusted for each subject. In a recent study, the authors have
shown how additional relaxation of accommodation can be
achieved using +2.00 D fogging lenses for central refraction
using as well an open-field autorefractor.27 Descriptives (mean±SD) were obtained for the refraction vector
components M, J0 and J45 emerged from the Fourier expansion
of the refraction function, as recommended by Thibos.32
For those measurements obtained with the +2.00D
fogging lenses, this dioptric value was subtracted from the
spherical component of the refraction in clinical notation
before converting to vector components. This value is the
equivalent refraction in the ocular plane for a working vertex
distance of 12 mm. The SPSS statistical package v.15 (SPSS Inc., Chicago,
IL, USA) was used to conduct the statistical analysis. The
Kolmogorov-Smirnov Test was applied to evaluate the normality
of data distribution. The Kruskal-Wallis Test and ANOVA
were completed to evaluate if different conditions yielded statistically
different values of M, J0 and J45 for non-parametric
and parametric variables. When normality could not be assumed,
the Wilcoxon Signed Ranks Test was used to perform a paired comparison between techniques, whereas the Paired-
Samples Test was used for normally distributed variables. The central objective refraction obtained under cycloplegia
for the whole sample, GS_cycl, ranged between -9.00 D
and +2.25 D for the value of the sphere, with a maximum
astigmatism of -2.25 D. The mean value of the spherical
equivalent (M) refraction was -0.85±2.27 D (GS_cycl). A
total of 35.0% of the sample had myopia (M=-3.42±2.05D),
32.5% had emmetropia (M=+0.23±0.34D) and 32.5% had
hyperopia (M=+0.84±0.31D) according to the criteria defined
above, under the methods section. Table 1 presents the mean and standard deviation of the
M component obtained both for the entire sample as well as
for each of the three refractive groups, and under the different
measuring conditions (GS, GS_2D and GS_cycl). This table
also presents the statistical significance for comparisons among
different measuring techniques and eccentric position in the
nasal and temporal fields. All values found for GS_cycl and
GS_2D were more positive or less negative than those obtained
with GS (without cycloplegic or fogging lens). Statistically
significant differences across all techniques at different eccentric
points were found for hyperopes except at 20º in the
temporal field and for emmetropes in the temporal field and
20º in nasal fields (P<0.05, Kruskal-Wallis Test). No statistically
significant differences were found across techniques at each
of the eccentric points in myopes (P>0.05) although there is a clinically significant trend towards more negative values with
GS. Figure 1 graphically illustrates these differences. No statistically significant difference was found across
techniques neither at the center nor at eccentric locations
for the astigmatic components J45 and J0 (P>0.05, Kruskal-
Wallis Test), as illustrated in figures 2 and 3. Table 2 presents the differences corresponding to paired
comparisons between techniques for the three refractive groups
and at different eccentric locations. This table does not attempt
to compare M values between different refractive groups, which
of course, are expected to change according to refractive classification
of patients. Instead, this table provides information about
how the techniques compare to each other regarding refractive
status and eccentric location. In this sense, we observe that
there are statistically significant differences between those techniques
that aim for the relaxation of accommodation (GS_2D
and GS_cycl) and GS (P<0.001) but not between GS_2D
and GS_cycl, with the exception of the myopic group. Mean
differences between methodologies for different eccentric points
within each refractive group range from -0.02±0.41 for GS_cycl
minus GS_2D in the emetropic group to -0.54±0.50 for GS
minus GS_2D in the myopic group. None of the comparisons
considering eccentric location as a factor showed statistically
significant differences (P>0.05, Kruskal-Wallis Test). With the results of the present study we have showed that
relaxation of accommodation can be attained using fogging spectacle lenses, when peripheral measurements are carried
out with an open-field autorefractor. In a previous study we
had already shown that the same could be achieved for central
measurements,27 so in this sense, the present results could
be expected. However, after that study, we asked ourselves
if prismatic effects or oblique incidence of light could be a
source of error when peripheral measurements were to be
obtained through the fogging lens. Nevertheless, the present
study also showed that the influence of oblique incidence
of light, with potential effects on astigmatism or prismatic
effects, is negligible when using this methodology. This is
supported by the absence of differences in peripheral refraction
(particularly regarding the J0 and J45 components)
when using the fogging lenses, as compared with the use of
cycloplegic agent alone. The study further showed that this
effect can be achieved in emmetropes and hyperopic young
adults where the accommodative system is very active. Even though this work aimed to analyze peripheral
refraction under several measuring conditions using a sole
instrument for different refractive groups, when conducting
a first analysis of the results obtained, we verified that the
differences across techniques found in peripheral refraction
are similar to those observed for the central measurements.
As expected, this study also showed that peripheral refraction
is more astigmatic than central refraction and is asymmetric,
in the sense that the astigmatism is more accentuated on
the temporal retina than on the nasal retina. This horizontal
asymmetry is higher for hyperopes and emmetropes than for
myopes, which is in agreement with previous studies.4,19,21,22 When analyzing the various refractive components individually,
the component M demonstrated that peripheral
refraction on hyperopes and emmetropes was more myopic
in the temporal retina. Meanwhile, at the same point, for
myopes, peripheral refraction presented slightly less myopic or more hyperopic values than those found for central refraction.
At the nasal retina, and for all refractive groups, the
values obtained were more hyperopic or more positive than
those found in the central region. This trend was also verified
by Atchison et al. when analyzing the influence of age on
peripheral refraction using the same measuring technique
(GS). In a group of young adults (24 ± 3 years), for the point
20º in the temporal retina, they found more myopic values,
in approximately -0.70D for emmetropes and similar values
for an adult population of hyperopes by -0.80D (age 59 ± 3
years). They also found that there were small variations between
the nasal and the central region of the retina, with smaller
differences being observed in the myopic group, which is
in good agreement with our results.19 In another study conducted
by Seidemann et al. using a different autorefractor the
results were slightly different, as all refractive groups showed
a trend towards more myopic values in the periphery. Similarly to our work, in that study smaller changes were
found in the nasal retina for all refractive groups.4 Similar
values were also reported by Calver et al. and by Atchison
et al., who presented results for myopes and emmetropes
only.4,21,22 Peripheral refraction studies usually reveal higher values
of astigmatism on the peripheral retina than in the center,
reaching its highest values in the temporal retina. When
analyzing the astigmatism J0 separately, we also observed
an asymmetry between the temporal field and the central
retina, being most negative at the temporal side by -0.63D
for hyperopes, -0.68D for emmetropes and -0.47D for myopes.
Concerning the nasal retina, the refractive alteration, as
compared to the center, was less accentuated than for the
temporal retina (-0.14D for hyperopes, -0.09D for emmetropes
and +0.09D for myopes). These results are in agreement
with Atchison et al. for the three refractive groups.19 Calver et al. presented a curve and values very similar to the ones
found in the present study for the astigmatic component
J0.22 Similar results were also described by Atchison et al.
for the emmetropic group, and mainly for a myopic group
having refractive characteristics similar to those of the present
study. Concerning the results for the 20º region, they also
observed that the nasal retina presented a dispersion of values
in relation to the adjusted curve and more positive values of
J0, compared to the center.21 Astigmatism J45 varies in an almost linear way between
the nasal and the temporal retina, but with a smaller amplitude
for emmetropes (J45_20ºnasal – J45_20ºtemporal
= +0.05D). Values of this magnitude were also found by
Atchison et al., even if their tendency was the opposite. A
higher magnitude was found for the hyperopic and myopic
group (+0.15D). Regarding myopes, Atchison et al. found
amplitudes of approximately -0.25D, while Calver et al.
found higher values, of about 0.60D.21,22 Beyond the agreement with previous studies analyzing
similar eccentric areas of the visual field, the present results
showed that the same difference observed for the central
refraction between methods attempting to control the accommodative
response is maintained in the periphery, across the
horizontal meridian up to 20º. Regarding the particular case
of direct comparison between pharmacological and optical
relaxation of the accommodative response during refraction,
no statistically significant differences were found for the
three refractive groups between GS_cycl and GS_2D at all
eccentric locations along the horizontal meridian, with the
only exception of myopes, for whom the fogging lenses seem
to render higher accommodative control than cycloplegia.
At present we do not have an explanation for this different
behavior. In clinical terms, the fact that both accommodative
control strategies behave in an effective and similar way in
emmetropes and hyperopes is important, as a shift towards
myopia due to accommodation could misclassify a subject as
myopic or less hyperopic.33,34 One limitation of the study is the fact that using the same
power for the fogging lens, irrespective of the patient’s ammetropia,
could result in a slightly different accommodation
control. This seems to be particularly important in myopes, where higher differences between cycloplegic and fogging
lens scenarios seem to be present. However, using the same
lens allows us to keep the experimental conditions constant
across all refractive groups. In this study, we attempted to
proof that fogging lenses can be successfully used to perform
peripheral refraction, in the same manner it was to measure
central refraction,27 and this goal seems to be achievable with
no apparent influence on peripheral astigmatism when compared
with the no-fogging-lens situation. Therefore, the measurement of central and peripheral
refraction using fogging lenses to replace the use of cycloplegic
is valid in young adults, particularly in emmetropes
or hyperopes. Surprisingly, the use of fogging lenses doesn’t
have a significant impact on peripheral astigmatism because
of the oblique incidence of light on the fogging lens. This
is somewhat expected if we bear in mind that the light passes
through the lens perpendicularly, since only the eye is
turned. The potential impact of oblique incidence of light
on peripheral astigmatism could be more important if the
instrument were displaced or the head itself turned,35 as
opposed to the approach commonly used, where the subject
only turns their eyes. This could be at least in part attributed
to the small entrance pupil of this instrument.36 We cannot
ensure that the same results would be obtained when using
other instruments. The use of fogging lenses seems to provide an accommodative
relaxation comparable to that obtained with cycloplegic
in young-adults, without any measurable effect on peripheral
refraction related to oblique incidence or prismatic
effects. These results are valid for the Grand Seiko open-field
autorefractor when used to assess eccentric locations up to 20º in the nasal and temporal fields. The present results have
potential application in mass field studies involving population
where accommodative control is more critical and
cycloplegic administration less desirable, such as in children
screening, and particularly when conducted by non-medical
staff. However, cycloplegia has the additional advantage of
achieve pupil dilatation, thus allowing more peripheral data
to be obtained.
