A Quantitative Method for the Analysis of Bilateral Corneal
Asymmetry in Different Corneal Categories
Panagiotis Kompotiatis and Michael K. Smolek
ABSTRACT
PURPOSE: It is generally accepted that the presence of bilateral corneal
asymmetry correlates with certain corneal diseases and may be
the evidence of corneal abnormalities. A method is shown for the
quantification of bilateral corneal asymmetry in 5 different corneal
categories, making use of corneal topography data.
METHODS: Corneal topographies (Tomey TMS, Nagoya, Japan)
from both eyes of 74 subjects were reviewed by a cornea researcher
who classified them according to their characteristics into five
groups (normal, astigmatic, contact-lens-induced corneal warpage,
keratoconus suspects and bilateral keratoconus). Topography maps
were chosen from the clinical records of the LSU Eye Center. The
axial-diopters tables for each pair of corneas were extracted from their
corresponding corneal topographies. The resulting data from the
corneal topographies of each eye were interpolated at regular intervals
using a polar-coordinate system. Subsequently, the differences between
corresponding points of the axial-diopters tables of the fellow
eyes were calculated. The cumulative percentage of the resulting
values was calculated and its distribution was plotted for each pair of
eyes. A linear regression was fitted to the data of each cumulative distribution
plot. Finally, the slope of the linear regression of each pair of
eyes was calculated and statistical analysis was performed in order to
determine whether or not there is a statistically significant difference
between the slopes of each group of corneal categories.
RESULTS: A statistically significant difference (P<0.05) was found
between keratoconus and all the other groups. Furthermore, we
found a statistically significant difference between the normal and the
contact-lens-induced warpage group as well as between the normal
and the keratoconus-suspects group (P<0.05). There was no statistically
significant difference (P>0.05) neither between the normal and
the astigmatism group, nor between the astigmatism and the contact-lens-induced corneal warpage group, nor between the astigmatism
and the keratoconus-suspects group, nor between the keratoconussuspects
group and the contact-lens-induced corneal warpage one.
CONCLUSIONS: A method to quantify bilateral corneal asymmetry in
different corneal entities was devised. The lowest degree of bilateral
asymmetry was observed in the normal group, while the highest degree
of bilateral asymmetry was obtained for the keratoconus group. The
contact-lens-induced warpage group and the keratoconus-suspects
group presented with intermediate bilateral asymmetry.
(J Optom 2009;2:173-181 ©2009 Spanish Council of Optometry)
KEY WORDS: corneal topography; bilateral asymmetry; corneal warpage; astigmatism; keratoconus suspects.
RESUMEN
OBJETIVO: La teoría de que la presencia de una asimetría corneal
bilateral está correlacionada con ciertas enfermedades corneales y
puede ser la prueba de la existencia de anormalidades corneales está
ampliamente aceptada. Se presenta aquí un método para cuantificar
el grado de asimetría corneal bilateral para 5 tipos (categorías) de
corneas, utilizando para ello datos de topografía corneal.
MÉTODOS: Un investigador especialista en córnea analizó las topografías
corneales (Tomey TMS, Nagoya, Japón) realizadas anteriormente en
los dos ojos de 74 sujetos. A continuación, clasificó a dichos sujetos de
acuerdo a sus características corneales, asignándole a cada uno una de
estas cinco categorías: normal (NOR), astigmatismo (AST), moldeamiento
corneal producido por lentes de contacto (MCLC), sospecha
de queratocono (SQ) y queratocono bilateral (QB). Los mapas de
topografía corneal se cogieron de los historiales médicos de la clínica
oftalmológica LSU Eye Center. Para cada par de córneas se elaboró la
tabla de dioptrías axiales a partir de los datos de las topografías corneales
correspondientes. Para ello, los datos de las topografías corneales se interpolaron
a intervalos regulares (tanto en la dirección radial como en la
angular) utilizando un sistema de coordenadas polares. A continuación,
se calculó la diferencia (de potencia, en dioptrías) entre cada par de puntos
análogos, a partir de las tablas de dioptrías axiales correspondientes
a los dos ojos de un mismo sujeto. Para cada par de ojos, se calculó el
porcentaje acumulado de los valores resultantes y se representó gráficamente
esta distribución. Para cada curva de porcentajes acumulados, se
realizó un ajuste lineal de los datos. Finalmente, para cada par de ojos
se calculó la pendiente de esta recta de ajuste. Luego se llevó a cabo un
análisis estadístico de todos los datos para determinar si existe una diferencia
estadísticamente significativa entre los distintos tipos (categorías)
de córnea en lo que respecta al valor de dicha pendiente.
RESULTADOS: Se halló una diferencia significativa (P<0,05) entre
QB y el resto de las categorías. Además, hallamos una diferencia
estadísticamente significativa entre el grupo NOR y el MCLC, así
como entre el grupo NOR y el SQ (P<0,05). No se obtuvo una
diferencia estadísticamente significativa (P>0,05) ni entre el grupo
NOR y el AST, ni entre el grupo AST y el MCLC, ni entre el grupo
AST y el SQ, ni entre el grupo SQ y el MCLC.
CONCLUSIONES: Se ha ideado un método para cuantificar el grado
de asimetría corneal bilateral para diversos tipos de córneas. El
menor grado de asimetría bilateral se observó para el grupo NOR,
mientras que el grado más elevado de asimetría bilateral se obtuvo
para el grupo QB. Para los grupos MCLC y SQ se encontró un
grado intermedio de asimetría bilateral.
(J Optom 2009;2:173-181 ©2009 Consejo General de Colegios de
Ópticos-Optometristas de España)
PALABRAS CLAVE: topografía corneal, asimetría bilateral, moldeamiento corneal, astigmatismo, sospecha de queratocono.
From the Computer-Aided Diagnostics and Physiological Optics Laboratories, Department of Ophthalmology, Louisiana State University Health Sciences Center, New Orleans, Louisiana, USA.
Acknowledgements: Supported by Research to Prevent Blindness, Inc. A
quantitative method for the analysis of bilateral corneal asymmetry in different
corneal categories. This research was supported by a research grant NIH
R01EY14162 (MKS).
Financial disclosure: The authors have no commercial or financial interest
in the products or methods described in this manuscript.
Received: 4 July 2009
Revised: 22 October 2009
Accepted: 24 October 2009
Corresponding author: Michael Smolek. 2020 Gravier Street, Suite B
New Orleans, LA 70112 (USA)
e-mail: msmole@lsuhsc.edu
It is generally accepted that humans, as well as all other vertebrates, are characterized by external bilateral symmetry along the dorsoventral (DV) and anteroposterior (AP) axes. This is achieved through a variety of mechanisms, including a complex network of signaling mechanisms and gene regulation.1 On the other hand, it should be mentioned that internally, humans present significant bilateral asymmetry, which is mainly revealed in the development of the internal organs such as the heart and the digestive tract. This process begins early in vertebrate development and it includes the encoding of several regulatory proteins and cell signaling molecules.2
In the human cornea, it has been demonstrated that the corneal topographies of the same subject present a high degree of mirror-image symmetry (enantiomorphism).3 In the same study it was demonstrated that although the corneal topographies of the same subject were similar, there was a variation in the morphology across the corneal topographies of different subjects. According to another study involving 50,995 subjects, the corneal astigmatic axis of the subjects presented mirror-image symmetry rather than direct symmetry.4 This mirror-image symmetry of the astigmatic axis was observed for all types of astigmatism (i.e., cylinder power ≤1D, cylinder power >1D, with-the-rule astigmatism, against–the-rule astigmatism and oblique astigmatism).
Another significant category of corneal morphology is contact-lens-induced corneal warpage. It has been documented that it can be caused by all types of contact lenses, although it is believed to be more commonly caused by rigid contact lenses and, especially, those made of PMMA (polymethylmethacrylate).5 The corneal topography in contact-lens-induced corneal warpage shows specific characteristics, like a central irregular pattern, flattening underneath the decentered lens and peripheral steepening opposite the resting position of the decentered lens. The latter characteristic can cause a high degree of morphological resemblance to the topographic characteristics of early keratoconus, especially when the contact lens rides on the upper part of the cornea.6,7 Research aiming to differentiate between contact-lens-induced corneal warpage and keratoconus topographies may prove beneficial in clinical screening.8,9
Corneal dystrophies often cause a significant distortion of the cornea. The most characteristic disease causing prominent corneal paramorphosis is keratoconus. Keratoconus is a non-inflammatory disease in which the cornea progressively develops a conical shape because of thinning and protrusion. It usually appears in the early adulthood, causing with visual blurring and distortion. The most common finding during eye examination is highly irregular corneal astigmatism.10,11 Early keratoconus may be present without any noticeable clinical signs and, for this reason, corneal topography has proven to be very useful in the detection of corneal irregularities.12 Keratoconus is, in most cases, a bilateral disease, where one eye is initially affected and the other eye follows after a certain period of time.13 It has also been suggested that all the unilateral cases of keratoconus, if they are observed for a sufficient amount of time, will eventually progress to a bilateral disease.14
It has been shown that there is a significant asymmetry between fellow eyes in bilateral keratoconus. In one study the asymmetry was assessed using 13 unilateral quantitative topographical indices as well as the best spectacle-corrected visual acuity (BSCVA). The results of the comparison of these parameters between fellow eyes revealed a significant degree of asymmetry in the bilateral cases of keratoconus.15 Another extended study that was based not on topographical indices but on the measurement of several clinical parameters concluded that keratoconus is an asymmetrical bilateral disease, both in an absolute scale and also in comparison with normal, myopic contact-lens-wearing patients.16
Regarding the detection of keratoconus suspects and early keratoconus cases, corneal topography has proven to be very useful; in this context, many quantitative indices17,18,19,20,21 that contribute to the detection of corneal irregularities have been developed. There are relevant indices that assess unilateral corneal asymmetry, such as the I-S value, which estimates the asymmetry between the inferior and the superior parts of the cornea,17 the Surface-Asymmetry Index (SAI),18 which expresses the differences in corneal power between corresponding points of the same cornea that are 180º apart on the photokeratoscope mires, and the Opposite-Surface Index (OSI),19 which represents the greatest difference in average power between opposite sectors of a given cornea.
It is a common notion that while normal corneas show a high degree of bilateral symmetry, the same thing does not happen for corneas that develop irregularities, either as a result of a disease like keratoconus or because of a mechanical distortion like that caused by contact-lens-induced corneal warpage. The aim of this study was to assess bilateral asymmetry in five distinct categories of corneal topographies (normal, astigmatic, contact-lens-induced corneal warpage, keratoconus suspects, bilateral keratoconus) and to develop a quantitative method to express the degree of asymmetry.
METHODSThis is a retrospective study of the clinical data from 148
eyes (74 subjects). All data were collected from de-identified
historical patient data of the Louisiana State University
(LSU) Eye Center clinic (Baton Rouge, LA, USA). Corneal
topography was originally collected once the patient had read
and signed an informed consent. The corneal topographies
of both eyes of 74 subjects where reviewed by a corneal
topography researcher with the fixed Smolek/Klyce scale (30
to 67.5 in 1.5-steps). The
classification was done according
to the topographic characteristics of each pair of eyes; each
subject was assigned to one of the five groups of interest: the
1st group included 14 subjects that were classified as normal
(NRM, Cylinder power ≤1.5 D), the 2nd group included 11
subjects that presented astigmatism (AST, Cylinder power
>1.5 D), the 3rd group included 15 subjects that had contact-lens-induced corneal warpage (CLW), the 4th included 20
subjects that were classified as keratoconus suspects (KCS)
and the 5th included 14 subjects that suffered from bilateral
keratoconus (KC).
The AST group included 9 subjects that had with-the-rule astigmatism and 2 subjects that had against-the-rule astigmatism. The KCS included subjects whosecorneal topography exhibited an area of local steepening or irregular astigmatism (asymmetric bow-tie pattern) either in one or in both eyes. The KCS group also included subjects that didn’t have the afore-mentioned features but that exhibited, nevertheless, a type of bilateral asymmetry in their corneal topographies that didn’t follow a characteristic pattern. The KC included 14 subjects. For 7 of them, both eyes’ keratoconus were in the same stage of the disease (the mild stage for 2 subjects, and the moderate stage for 5 subjects). Regarding the other 7 subjects, the stage of the disease was different between fellow eyes: one eye had mild-stage keratoconus whereas the other had moderate-stage keratoconus.
The corneal topographies of the left eyes of all the subjects were flipped about their vertical axis with the help of a customized computer program that transposes left-eye data to right-eye data and vice-versa. This was done in order to correct for the naturally occurring mirror-image symmetry of fellow corneas. By means of the CTView program (Sarver and Associates, Inc, Carbondale, Il., USA) the corneal topography of each eye was sampled, based on a regular polar-coordinate system. The measurement points were distributed along radii, traced at 10˚-intervals. The distance between adjacent points on the same radius was 0.5 mm, with a maximum distance from the center of 5 mm. These data were presented on an axial diopters table with polar coordinates. The table consisted of 11 columns and 37 rows, with a total of 407 cells. The columns represented the point’s radial location (at steps of 0.5 mm on each radius, from 0 to 5 mm) and the rows its angular location (at steps of 10º for each mire, from 0º to 360º).
The axial diopters table for the right eye and the flipped left eye of each subject were exported to a Microsoft Excel spreadsheet and the differences between the values of corresponding points in the right and left corneas were then calculated, leading to the so-called “difference table”. When in one of the tables (left-eyeor right-eye) one of the values was missing, the corresponding cell in the final difference table was left blank. Subsequently, for each pair of eyes, the cell values of the final difference table were inputted into a statistical program (Minitab, State College, PA, USA). In Minitab the columns of each difference table were stacked to only one column and the cumulative percents of thevalues of that column were calculated. As a result, the final data for each pair of eyes consisted of two columns: the first one contained the values (in diopters) of the final difference tablen ascending order and the second column the corresponding cumulative percent value, which is calculated as the percentage of data points whose difference values are equal or below the value of that particular sample data point. For instance, a value of 100% means that all the other sampled points had a difference value below this one. The cumulative-percent values ranged from 0% to 100%. Finally, a linear regression model was applied tothe data of these two columns, using the first column (values of the difference table) as x-values and the second column (cumulative percents) as y-values. The graph in figure 1 shows the cumulative percentage distribution plots (with the corresponding linear-regression fit) for two pairs of eyes, one belonging to the normal group and the other to the keratoconus group. The equation of each linear regression is also shown; the slope of each of these lines is underlined.
In fact, the slope of each regression line was used as a
measure of the dispersion of the values of the final difference
table for each pair of eyes. Higher slope values corresponded
to low data dispersion, which was more prominent in the
normal group, while lower slope values corresponded to a
higher data dispersion, a pattern that was more typicalof the
keratoconus group. Table 1 shows the slope of the regression
lines for each one of the 74 pairs of eyes under study, classified
into the 5 different groups (NRM, AST, CLW, KCS and
KC). Figure 2 is the interval plot, depicting, for each of the 5
eyecategories under analysis, the slope’s mean value together
with standard error bars.
Finally, a One-Way Analysis of Variance test was conducted with SigmaStat (Systat Software, Inc, San Jose, CA, USA) in order to determine whether there was a statistically significant difference between the values of the slope across the different corneal categories. The normality test failed and a Kruskal-Wallis One-Way Analysis of Variance on Ranks test was performed. The analysis was concluded with Dunn’s method for all pairwise multiple comparison of the 5 groups.
RESULTSAccording to the Kruskal-Wallis test statistic value (H=42.238 with 4 degrees of freedom) there was a statistically significant difference between the 5 groups (P<0.001). The results of the test are shown in table 2. Dunn’s test was performed to determine which pairwise comparisons on ranks were statistically different. The Q test statistic value that was estimated according to Dunn’s method for each pair of groups indicated that there was a statistically significant difference (P<0.05) between KC and each of the other groups. Additionally, a statistically significant difference (P<0.05) was found between the NRM group and the KCS group as well as between the NRM and the CLW group. There was no statistically significant difference (P>0.05) between the NRM and the AST group or between the AST and the KCS group. It should be mentioned that, according to Dunn’s method, a comparison between the AST and the CLW warpage group or between the KCS and the CLW group was not performed because the test failed to detect a significant difference between the AST and the KCS groups. The results of Dunn’s test are shown in table 3.
DISCUSSIONIn this study, a quantitative method was developed in
order to assess the bilateral asymmetry for five different
corneal categories, without making use of unilateral cornealtopography
quantitative indices, which only describe the
corneal characteristics of individual eyes.15,17,18,19,20,21 The comparison
of fellow eyes was based only on the raw data that were
extracted from the corneal topographies of each eye by means
of the CTView application. The data for each eye were interpolated at regular intervals, according to a polar-coordinates
system. For each pair of matching points (one for the left eye,
another for the fellow right eye) having the same coordinates,
the difference in axial diopters was calculated. The concept
was that if there is a significant steepening (higher power)
or flattening (lower power) in an area of one of the fellow
eyes, then that would result in a larger diopter difference
with respect to the matching point (i.e., that with the same
coordinates) in the fellow eye.
The purpose of using raw data was to avoid devising a method based solely on proprietary indices. The current method can be used on any topography system.
The statistical analysis was based on the resulting table
containing the differences between the data of fellow eyes.
For each pair of eyes, the raw data of the difference table were
tested with the Kolmogorov-Smirnov normality test, which
failed for almost all pairs of eyes within all the groups. This
suggested that the data for each pair of eyes did not follow a
normal distribution and, as a result, it
could not be described
by means of the related statistical parameters such as the
mean and the standard deviation. Another alternative could
be the use of the median, which was rejected because it is not
representative of the dispersion of the data. The interquartile
range (75th percentile – 25th percentile) was also used but
the results were not as significant as those obtained with the
slope-analysis method. According to this latter method, the
slope of the linear regression of the cumulative-percentage
distribution plot describes the dispersion of all the data, for
each pair of eyes. A high slope value means that the data
show a small dispersion, which corresponds to a low degree of bilateral asymmetry, while lower values of the slope means
that the data show a higher dispersion, which corresponds to
a higher degree of bilateral asymmetry.
For all the subjects within the normal group, the resulting slope values were very high (>100). This was to be expected, as the mirror-image symmetry of fellow normal corneas has been demonstrated previously.3 A characteristic example of normal fellow corneas is shown in figure 3. The value of the slope for this example was 161.9. All the pairs of eyes within this category were characterized by visually apparent mirror-image symmetry, which also became evident during the analysis of the data. Conversely, the AST group had slopes that presented a significant variance, although the visually apparent mirror-image symmetry of the astigmatic axis was prominent for all the subjects within this group4. The variance of the resulting slope within this group may be attributed to wider range, in terms of morphology, of the corneal topographies of the fellow eyes, even though this is not readily noticed just by means of visual inspection. A characteristic example of corneal astigmatism is shown in figure 4. The value of the slope for this example was 61.8.
The results from the CLW group revealed a significant bilateral asymmetry. For 13 out of the 15 pairs of eyes within this group, the values of the slope were below 100, which were in agreement with the qualitative visual assessment of the corneal topographies, which already indicated a significant bilateral asymmetry. This suggests that the distortion of the cornea from contact lens usage causes, in most cases, a significant bilateral asymmetry. In figure 5, an example of bilateral asymmetry taken from this group is shown. The value of the slope, for this example, was 55.5. In this same group there were also two pairs of eyes that presented bilateral mirror-image symmetry. The value of the slope for these 2 pairs of eyes was above 100. The corneal topographies for one of these pairs of eyes, having a high value of slope (158), are shown in figure 6.
The KCS group also showed a moderately high variability in terms of slope of each pair of eyes, while the mean slope was low. Most subjects presented bilateral asymmetry and the slope, in most cases, did not exceed 100. A characteristic example of bilateral asymmetry in KCS is shown in figure 7; in this particular case, the slope was 23.48. There were some cases that, despite a visually apparent abnormal morphology in both eyes, the bilateral asymmetry turned out to be very low and, as a result, the corresponding slope value was high. An example of such a case is shown in figure 8; the slope in this case was 122.6.
The KC group showed the lowest slope values of all groups. When the disease was in a different stage in one eye in relative to its fellow eye, then the value of the slope was even lower than when the disease was in the same stage in both eyes. An example of a bilateral case of KC showing different stages of the disease in each fellow eye is shown in figure 9. The presence of asymmetry is prominent, and it is sufficiently represented by the very low value of the slope for each pair of eyes (13.1). figure 10 depicts another case of bilateral keratoconus, where the stage of the disease is the same in both eyes. The degree of asymmetry is lower than in the previous example (see figure 9); however, the value of the slope is still significantly low (57.5).
Statistically, the KC group was well differentiated from all the other groups; its high degree of bilateral asymmetry characterizes this disease. Regarding the other corneal categories, the KCS and the CLW groups are statistically different from the normal group, but not from each other. This indicates that in both groups exists bilateral asymmetry but, unfortunately, it is not possible to differentiate between them with this method alone. Furthermore, it is worth noting that given the 95% confidence interval for the mean of the slope of the normal group, 18 out of the 20 KCS and all the bilateral KC cases would be classified as non-normal with this method. Figure 11 contains a plot that represents the individual slope values for each pair of eyes in each group and it also includes interval bars that indicate the 95% confidence interval for the mean of the slope value within each category. However, given the limited size of the sample, the interpretation of these results should be made with particular caution.
CONCLUSSIONSIn this study we have presented a quantitative method that describes bilateral asymmetry in different corneal categories. This method was designed to work with any corneal topography system that returns axial-curvature values in diopters. It does not use proprietary indices specific to one system. The method was highly effective to differentiate normal and astigmatic corneas from abnormal conditions related to ectasia and corneal warpage. However, it wasn’t able to differentiate CLW from KCS. Nevertheless, the method may prove useful to detect when a warped cornea returns to its normal state. This method can be complementary to all the screening procedures that are used nowadays based on the description of the characteristics of the cornea. It provides additional information that will help to detect diseases, like KC, when they are still in its early stages. Moreover, it will contribute towards a more efficient screening of refractive surgery candidates, because it evaluates both corneas, rather than evaluating each cornea separately.